XENON KRYPTON DETERMINATOR – KX 2001 


Glove Box Model

Special Features:
  1. Calibration using Dosing gas (100%) Xe and Kr in system test mode saves precious and costly calibration standards. Alternatively, 1% Kr/Xe Gas Mix can also be used.
  2. All timings for the various modes can be controlled through the membrane keypad. The KX PPC software also allows for this flexibility in case of the PC control option.
  3. Step wise Fault diagnosis on the operator panel console.
  4. Multiple safety interlocks for electrode Furnace.
  5. Removable lower and upper inserts offer ease in replacement and wide sample weight range by changing crucibles.
  6. Current Mode, Time for Mode completion etc. Displayed on the LCD of the OP console as well as KX PPC software.
  7. On line help, Method Development data as well as user manual can be accessed on line using OP console as well as the KX PPC software.
  8. Temperatures upto 3400° C using EF-2002-MT Electrode Furnace. 
Specifications:

Range  
Krypton 1.0-5000 PPM at 1 gm sample  
Xenon   1.0-5000 PPM at 1 gm sample

Accuracy  
< 2% of the reading

Sensitivity  
1.0 ppm for Xe/Kr

Gas Required  
Carrier Gas : Helium 99.999% pure Calibration  : Cal. Gas or Standard gas Pneumatic Gas: Compressed Air / N2, Gas must be water and oil free

Sample Weight  
Range      : 0.50 To 5.000 grams

Analysis time (including out gas)  
Total Time     : 20 Min Nominal

Detection Method  
Macro High Flow TCD for Kr and Xe

Sample Trap  
U-Type SS trap filled with Absorbent material Heated with 2KW transformer with control using CAL 9400. Collection at –60° C and heated at 400° C

Chemical Reagents  
Anhydrous Magnesium Perchlorate  
(MgCIO4), Sodium Hydroxide on an inert base, Cu Turnings